Features & Specifications
  • Easy-view, touch-panel, digital display with soft key operation
  • Intuitive design allows for easy set-up and operation
  • Compatibility with any HPLC system
  • Low-vacuum nebulizer minimizes mobile phase out gassing
  • Response unaffected by mobile phase pH
  • Small footprint, quiet operation
    Specifications - Model NQAD

    Inlet mobile phase flow rate:  

    0.1 – 1.2 mL/min

    Inlet connection:           

    Standard 10-32 receptacle

    Evaporator temperature:          

    35° to 100° C or off, programmable

    Wetted materials:         

    PTFE, PFA, 316 SS, PEEK, Ruby, Kalrez®

    Gas requirements:        

    Clean dry air or nitrogen
    <10 SCFH (4.7 L/min) at 40 PSI regulated (276 kPa)

    Condensing liquid:        

    Water

    User interface:             

    5” Color touch panel display
    Soft key menuing
    Real-time chromatograph
    Status and setup displays

    Back panel:     

    Water inlet (500ml bottle)
    Gas inlet ¼” Swagelok®
    Gas effluent outlet 3/8” Swagelok
    Liquid effluent outlet ¼-20 receptacle
    Analog/digital interface terminal connectors
    Power connector
    USB/Serial connectors

    Power requirements:    

    150 watts 100-240vac 50-60Hz

    Operating temperature:

    15 to 35º C

    Size:    

    6” x 17” x 12.5” (15.5cm x 43cm x 32cm) WDH

    Weight:

    20 lbs (9.1 kg)

    Analog output: 

    0 - 1.00 volt selectable gains x1 to x500

    Product safety/compliance       

    CE Certification based on 89/336/EEC, 72/23/EEC, EN55011, EN61326-1, EN61010-1

    To Order

    NQAD

    Nano Quantity Analyte Detector

    Options

    QT-5001

    IQ/OQ/PQ Documentation

    Specifications are subject to change without notice.
    Kalrez® is a registered trademark of DuPont Performance Elastomers. Swagelok® is a registered trademark of Swagelok Company. 

Front Panel