Forgot password?

NRM Features and Specifications

Model 8000 NRM  
Displayed Measurement Range 0.001 to 60 ppb
Response time to impurity change < 90 seconds
Sample flow rate 2 mL/min (calibrated drip counter flow meter)
Total flow rate 100 mL/min
Inlet water pressure 138 to 483 kPa; 20 to 70 psig (a suitable water drain is required)
Compressed air flow rate/pressure 5 L/min at 345 kPa; 50 psig (compressed air should be dry and free of particles. Nitrogen may be substituted for compressed air)
Residue drying temperature 120C; 248F
Wetted surface materials Sapphire, PFA Teflon, PTFE, and 316 L stainless steel
Detector working fluid Water, supplied by UPW system
Ambient temperature range 15-35C; 59-95 F
Maximum water temperature 50C; 122F
Dimensions (WDH) 43 x 36 x 22 cm; 17 x 14 x 8.5 inches
Weight 8 Kg; 18 lbs
Power Universal 100 to 230 VAC, 50/60 Hz, 125 VA
Output USB, RS-232 and 4-20 mA; Ethernet adapter included
Ultrapure water inlet 114 inch PFA Flaretek
Waste outlet 1/2 inch SS Swagelok
Compressed air inlet 1/4 inch SS Swagelok
Software Data acuisition program included, Windows XP platform